Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers
Original Article | Otorhinolaryngology: Ophthalmology & Optometry

Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers

Verónica Noya-Padin1,2 ORCID logo, Hugo Pena-Verdeal1,2 ORCID logo, Jacobo García-Queiruga1,2,3 ORCID logo, Noelia Nores-Palmas1 ORCID logo, Alba Castro-Giraldez1 ORCID logo, Eva Yebra-Pimentel1,2

1Departamento de Física Aplicada (Área de Optometría), Facultade de Óptica e Optometría, Universidade de Santiago de Compostela, Santiago de Compostela, Spain; 2Health Research Institute of Santiago de Compostela (IDIS), Santiago de Compostela, Spain; 3Facultad de Óptica y Optometría, Universidad de Santiago de Compostela, Santiago de Compostela, Spain

Contributions: (I) Conception and design: All authors; (II) Administrative support: H Pena-Verdeal, E Yebra-Pimentel; (III) Provision of study materials or patients: J García-Queiruga, H Pena-Verdeal, A Castro-Giraldez; (IV) Collection and assembly of data: J García-Queiruga, H Pena-Verdeal, A Castro-Giraldez; (V) Data analysis and interpretation: H Pena-Verdeal, E Yebra-Pimentel; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors.

Correspondence to: Jacobo Garcia-Queiruga, OD, MSc, PhD. Facultad de Óptica y Optometría, Universidad de Santiago de Compostela, Campus Vida s/n, 15782 Santiago de Compostela, Spain; Departamento de Física Aplicada (Área de Optometría), Facultade de Óptica e Optometría, Universidade de Santiago de Compostela, Santiago de Compostela, Spain; Health Research Institute of Santiago de Compostela (IDIS), Santiago de Compostela, Spain. Email: jacobogarcia.queiruga@usc.es.

Background: Precise quantification of corneal parameters enhances the clinical of the anterior corneal surface for various clinical assessments, with ongoing research focusing on optimizing everyday practice through device comparison studies. The study aimed to compare the agreement on the anterior corneal surface characteristics values measured by four commercially available devices: EyeSys 2000, Oculus Easygraph, ARK700A and ARK510A.

Methods: Observational cross-sectional study involving 50 eyes from healthy volunteer recruited among subjects attending for an eye examination. Corneal radius minimum keratometry reading (min-K) and maximum keratometry reading (max-K), astigmatism axis and eccentricity were measured in all participants with EyeSys, Oculus Easygraph, ARK700A and ARK510A. Measurements were performed three times only in the right eye in random order by four optometrists who were unaware of the results of the other researchers.

Results: A statistically significant difference occurred for the measurements of the corneal radius (Greenhouse-Geisser and Huynh-Feldt: P≤0.001 for both max-K and min-K radius) and the axis of astigmatism (Friedman: P<0.001), between the four devices. Pairwise comparisons revealed no radius differences between Oculus Easygraph and ARK510A (Sidak: min-K radius, P=0.231; max-K radius, P=0.224), and no axis differences in all pairwise analyses that have not imply the EyeSys (Wilcoxon: all P≥0.130). There was no statistical difference in the eccentricity value obtained by the topographers (paired t-test: P=0.507).

Conclusions: The present study showed that EyeSys 2000 measurements differed from those of the other devices across all parameters. Alternatively, both topographers showed a perfect agreement concerning eccentricity measurements in everyday practice.

Keywords: Agreement corneal topographers; agreement autokeratometers; EyeSys; Easygraph; ARK


Received: 29 December 2023; Accepted: 11 June 2024; Published online: 23 July 2024.

doi: 10.21037/amj-23-257


Highlight box

Key findings

• Agreement cross-sectional between devices on corneal surface measurements.

What is known and what is new?

• Differences in corneal radius and axis measurements between devices was found. Oculus Easygraph and ARK510A showed no difference in radius measurement.

What is the implication, and what should change now?

• EyeSys 2000 showed differences in axis measurements with the other devices. There was no difference in the eccentricity value obtained by the topographers.


Introduction

Precise quantification of corneal topography parameters provides a tool to define the optical characteristics of the anterior corneal surface to a proper contact lens fitting as well as a method for the early diagnosis and monitoring of corneal ecstatic disorders, accurate intraocular lens power calculations, and preparation of corneal refractive surgery (1-3). Keratometry is a technique that measures the central area of the anterior cornea where two axes 90° apart are determined. These two axes are the steepest and the flattest meridians of the central area of the anterior cornea. The steepest meridian has the highest radius of curvature, which fits with the maximum keratometry reading (max-K). On the other hand, the flattest meridian is the smallest radius of curvature, that fits with the minimum keratometry reading (min-K) and is perpendicular to the max-K (4).

Keratometric measurements can be acquired using either conventional keratometry or corneal topography (2,5-8). These devices provide clinicians with the characteristics of the anterior surface of the cornea based on the two principal radii (min-K and max-K), the astigmatism axis and the shape factor (2). The anterior corneal curvature radius determines two-thirds of the optical refractive power of the eye and is thus a relevant parameter of the ocular optical system (9). The shape factor of the cornea can be described in terms of eccentricity based on the mathematical description of an ellipse from 0.00 (a circle with no flattening) to 1.00 (maximum flattening in the periphery) (10); this parameter is usually only provided by the topographers, however, the problem with these instruments is that their sophistication inevitably induces a penalty in terms of cost and the induction of other errors because of the capture systems used, while an assessment of corneal topography using a simple instrument like keratometers provides the possibility of topographical assessment for the clinician with access only to basic instrumentation.

The differences of the precision and comparison between devices are still a relevant research topic for the optimization of the everyday practice (1,11-13). The present study aimed to compare the agreement of the anterior corneal surface characteristics values measured by four commercially available devices (two corneal topographers and two autokeratometers): the EyeSys 2000 (EyeSys Vision, United States), the Oculus Easygraph (Oculus Optikgeräte GmbH, Germany), the ARK700 (Nidek Technologies, Italy), and ARK510A (Nidek Technologies).


Methods

Study design

Measurements were performed in random order using four devices, two corneal topographers and two autokeratometers, following manufacturer’s instructions: (I) EyeSys Corneal Analysis System 2000, (II) Easygraph, (III) ARK700, and (IV) ARK510A (5-8). Corneal radius (min-K and max-K) and astigmatism axis were measured by all the devices, while the eccentricity values were measured only with the Oculus Easygraph and the EyeSys 2000 topographers. All measurements were performed three times on each device only in the right eye of each subject to avoid overstating the precision of statistical estimates by four expert optometrists who were not aware of the results of the other researchers (14).

Sample

Fifty volunteer participants (mean age =22.9±5.85 years; range, 19–29 years) were recruited among students and patients attending the centre for an eye examination. Participants were excluded if they had a history of conjunctival, scleral, or corneal disease, prior eye surgery, dry eye disease, meibomian gland dysfunction, glaucoma, diabetes mellitus, a thyroid disorder, wore contact lenses, were pregnant or breast feeding (15,16). Participants were only included if their refractive spherical error was less than ±6.00 D and/or astigmatism was less than 4.00 D (17). No one participant was under any type of topical or systemic medications or used artificial tears at the time of the study that could influence the measurements. To eliminate any diurnal variation in corneal curvature or tear physiology, all measurements were made in the morning at the same hour between 10:00 and 12:00 am (6,18).

Statistical analysis

SPSS statistical software v.23.0 for Windows (SPSS Inc., United States) was used for data analysis. Significance was set at a P≤0.05 for all statistical tests (19). Before analysis, the normal distribution of the data was checked using the Kolmogorov-Smirnov test.

Radius values showed a normal distribution (Kolmogorov-Smirnov test: P≥0.074 for all the examined values), therefore differences between measurements on devices were assessed using Greenhouse-Geisser or Huynh-Feldt correction based on Mauchly’s W test of sphericity, while pairwise analysis was performed using the Sidak post-hoc test (20). Since the axis showed a non-continuous distribution (Kolmogorov-Smirnov test: P≤0.001 for all the examined values), differences in this parameter between devices were assessed using the Friedman test, while the Wilcoxon test was used to detect significant pairwise differences; to avoid type I errors arising from multiple comparisons in the axis values, statistical significance for the Wilcoxon test to be divided by the number of comparisons performed to give a P≤0.0083 (Bonferroni adjustment) (21).

Eccentricity values showed a normal distribution (Kolmogorov-Smirnov test: all P≥0.194). For eccentricity agreement between topographers, Bland-Altman’s procedures were used. This method describes the correlation or similarity between two variables, representing averages versus differences (19,22). Thus, differences between the sets of measurements obtained in the two sessions were examined. Differences were assessed using a paired t-test for related samples and 95% limits of repeatability were also calculated [mean difference ± 1.96 × standard deviation (SD) differences] (19); 95% limits of agreement (95% LoAs) were also calculated (mean difference ± 1.96 × SD), as well as the exact 95% confidence intervals (95% CIs) for upper and lower LoAs considered as a pair (mean difference ± ct0.025 × SD; mean difference ± ct0.975 × SD) (19,23).

Ethical consideration

The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013). The study was approved by the Ethics Committee of Universidade de Santiago de Compostela (USC-40/2020) and informed consent was taken from all individual participants.


Results

There was a statistical difference in the measurements of both corneal radii, max-K [Mauchly’s W test: P=0.001, epsilon (ε) =0.561; Huynh-Feldt correction: P<0.001, Table 1] and min-K (Mauchly’s W test: P<0.006, ε =0.847; Greenhouse-Geisser correction: P<0.001, Table 1), between the four devices. Pairwise comparisons showed that this difference was not true when comparing Oculus Easygraph topographer and ARK510A autokeratometers (Sidak test: min-K radius, P=0.23; max-K radius, P=0.22, Table 2), whereas there was a statistically significant difference for the remaining comparisons (Sidak test: all comparisons max-K or min-K radius, P≤0.02, Table 2).

Table 1

Descriptive statistics and difference analysis between the flat or the steeper radius values obtained in each device

Parameter Device Mean/median SD/IQR P value
max-K (mm) EyeSys 2000 7.81 0.26 <0.001
Oculus Easygraph 7.72 0.26
ARK700 7.74 0.26
ARK510A 7.70 0.28
min-K (mm) EyeSys 2000 7.93 0.30 <0.001
Oculus Easygraph 7.87 0.29
ARK700 7.88 0.30
ARK510A 7.86 0.30
Astigmatism axis (°) EyeSys 2000 159 131–168 <0.001§
Oculus Easygraph 175 160–180
ARK700 176 144–1
ARK510A 176 147–180

N=50. Mean and SD displayed for parametric variables while median an IQR for non-parametric variables. , Huynh-Feldt correction; , Greenhouse-Geisser correction; §, Friedman test. SD, standard deviation; IQR, interquartile range; max-K, maximum keratometry reading; min-K, minimum keratometry reading.

Table 2

Analysis of differences of radios and axis results obtained in each device

Parameter Pair Mean difference ± SD P value 95% LoAs
Minimum (95% CI) Maximum (95% CI)
max-K (mm) EyeSys 2000-Oculus Easygraph 0.06±0.034 <0.001 −0.006 (−0.023, 0.004) 0.125 (0.115, 0.143)
EyeSys 2000-ARK700 0.05±0.036 <0.001 −0.026 (−0.044, −0.015) 0.116 (0.105, 0.134)
EyeSys 2000-ARK510A 0.06±0.032 <0.001 0.002 (−0.015, 0.012) 0.126 (0.116, 0.143)
Oculus Easygraph-ARK700 −0.01±0.033 0.003 −0.078 (−0.095, −0.068) 0.049 (0.039, 0.066)
Oculus Easygraph-ARK510A 0.004±0.026 0.23 −0.046 (−0.059, 0.038) 0.055 (0.047, 0.068)
ARK700-ARK510A 0.02±0.023 <0.001 −0.026 (−0.038, −0.019) 0.064 (0.057, 0.076)
min-K (mm) EyeSys 2000-Oculus Easygraph 0.09±0.047 <0.001 −0.003 (−0.027, 0.011) 0.182 (0.167, 0.206)
EyeSys 2000-ARK700 0.08±0.040 <0.001 −0.002 (−0.022, 0.010) 0.154 (0.142, 0.174)
EyeSys 2000-ARK510A 0.10±0.090 <0.001 −0.073 (−0.119, −0.045) 0.282 (0.254, 0.328)
Oculus Easygraph-ARK700 −0.01±0.038 0.02 −0.091 (−0.111, −0.079) 0.064 (0.052, 0.085)
Oculus Easygraph-ARK510A 0.02±0.087 0.22 −0.156 (−0.200, −0.129) 0.186 (0.159, 0.231)
ARK700-ARK510A 0.03±0.085 0.02 −0.138 (−0.181, −0.112) 0.195 (0.169, 0.239)
Astigmatism axis (°) EyeSys 2000-Oculus Easygraph −14.4±14.69 <0.001 −43.16 (−50.66, −38.62) 14.42 (9.88, 21.92)
EyeSys 2000-ARK700 −10.5±14.60 <0.001 −39.11 (−46.56, −34.59) 18.11 (13.59, 25.55)
EyeSys 2000-ARK510A −11.7±1.68 <0.001 −32.63 (−38.08, −29.33) 9.23 (5.93, 14.68)
Oculus Easygraph-ARK700 3.9±2.25 0.61 −35.81 (−46.15, −29.55) 43.55 (37.29, 53.89)
Oculus Easygraph-ARK510A 2.7±11.93 0.13 −2.71 (−26.80, −17.02) 26.05 (22.36, 32.14)
ARK700-ARK510A −1.2±12.50 0.26 −25.71 (−32.09, −21.84) 23.31 (19.44, 29.69)

N=50. , Sidak test; , Wilcoxon test. SD, standard deviation; LoAs, limits of agreement; CI, confidence interval; max-K, maximum keratometry reading; min-K, minimum keratometry reading.

There was a statistical difference in the measurements of the astigmatism axis between the four devices (Friedman test: P<0.001). The pairwise analysis showed that the EyeSys 2000 results were statistically significantly different from the other devices (Wilcoxon test: all P≤0.001, Table 2), while all the other devices showed no paired differences between them (Wilcoxon test: all P≥0.13, Table 2).

There was no statistical difference in the eccentricity value obtained by the two topographers (paired t-test: P=0.51, Table 3 and Figure 1).

Table 3

Descriptive statistics and paired t-test between the eccentricity results obtained in each topographer

Device Mean ± SD Mean difference ± SD P value 95% LoAs
Minimum (95% CI) Maximum (95% CI)
EyeSys 2000 0.458±0.083 −0.009±0.097 0.51 −0.200 (−0.250, −0.170) 0.182 (0.151, 0.231)
Oculus Easygraph 0.467±0.130

N=50. SD, standard deviation; LoAs, limits of agreement; CI, confidence interval.

Figure 1 Mean versus differences (Bland-Altman plot) between the eccentricity values obtained by the two devices in 50 participants. The thick solid horizontal line indicates the mean difference while the thin solid horizontal lines the 95% LoAs (mean difference ± 1.96 × SD). The dashed horizontal lines indicate the 95% CI of the LoAs. LoAs, limits of agreement; SD, standard deviation; CI, confidence interval.

Discussion

In clinical practice, accurate corneal topography measurements are essential for contact lens fitting or the detecting of significant corneal shape alterations. The measurement of the anterior corneal curvature or keratometry could be performed with a variety of devices.

Considering the mean difference and the 95% LoAs for max-K or the min-K measurements, the worst agreement was obtained between the EyeSys 2000 against the other devices; this device overestimated the radius values of the other devices. Inter-device differences, although small (near or over 0.10–0.20 mm), were suggested to be clinically significant for certain applications such as contact lens fitting and follow-up examinations. The same situation could be observed on the astigmatism axis measurement, where the EyeSys 2000 device may generate differences with the other devices higher than 10°, which could highly compromise a decision or evaluation of the initial contact lens chosen during a fitting procedure. From the studied devices here, the EyeSys 2000 is one of the most assessed for accuracy or repeatability in the literature (5,24-26). Previous studies of the EyeSys 2000 showed acceptable repeatability or agreement with other devices in the measurement of curvature parameters on human corneas or calibrated reference standards (5,25,27), while higher disparity was obtained on the astigmatism axis assessment than other devices or in the accuracy while the measured surface become more min-K flattening (5,26,27).

To the authors’ knowledge, whereas a statistically significant difference occurred with the Oculus Easygraph, this device was used in previous research on corneal optical quality and keratometry assessment (7,8). In the present study, a lower value of mean difference min-K, max-K, or the axis when the readings were compared to the autorefractometer results. However, while most of the 95% LoAs seem to be narrow, a few of them showed a bias that could compromise a clinical setting (e.g., min-K: Oculus Easygraph-ARK510A, 95% LoAs =−0.156, 0.186; axis: Oculus Easygraph-ARK700, 95% LoAs =−35.81, 43.55). It was previously stated those differences could be attributed to the measurement area of each device such as the normal prolate shape of the cornea is max-K in the centre and higher K-readings should be expected with smaller measurement areas (4).

Both studies autokeratometers studied showed a reliable agreement between them when min-K or axis was compared, whereas differences were found regarding the max-K measurements where ARK700 often overestimated the values in reference to ARK510A readings; nevertheless, the mean differences found could be considered non-clinically relevant (max-K: ARK700-ARK510A, mean difference ± SD =0.02±0.023 mm). Other autokeratometers models were also studied by other authors, showing good repeatability and agreement when compared with other automated devices and manual Javal keratometers (4). On the other hand, in concordance with present results other studies have proposed that EyeSys 2000 and the automated ARK700A keratometer lack interchangeability in their results (6).

Eccentricity could be defined as the flattening rate of the central curve radius of a surface and is a way to quantify aspheric changes across the diameter of that curve (10); the higher the value, the more quickly the surface flattens from the centre to the periphery (10). This variable has a relevant influence on the gas permeable fitting and evaluation process (e.g., irregular corneas or ortho-k fitting) (28). In the present study, EyeSys 2000 and Oculus Easygraph showed a high concordance in the readings of this parameter (eccentricity: EyeSys 2000-Oculus Easygraph, mean difference ± SD =−0.009±0.097). A few researchers have studied the agreement in the eccentricity values between different commercially available topographers, where significant differences among them were found (29).

It has been hypothesized that the differences in the readings of all those parameters between devices could be most likely due to the different imaging methods used by each device (29,30); topographers are based on an image capture of the Placido disc reflection from the anterior cornea while keratometry is based on the protestation and size measurement of an image in the cornea assumed as a convex mirror (29,30). In addition, it has been proposed that differences between keratometry assessments could emerge from not consistent use of an index of refraction or measurement area (the closer to the centre, the max-K the cornea will appear) by manufacturers (4). On the other hand, the differences could emerge from the capture times (30); topographers require a longer capture time to acquire multiple images or to focus the image tan autokeratometers, therefore factors such as the fixation of the subject, compensatory saccadic movements, and ability to keep the eye open may influence on the measurements.

The present results are limited by the sample used, which despite the fact it was relatively larger (50 participants), was only composed of healthy participants. Future studies on those devices should include a second group composed of a well-designed group with a specific corneal topographic disruption condition (e.g., Keratoconus or post-Lasik patients) in order to enhance the applications of the finding.


Conclusions

As previously indicated, the findings of this study demonstrate that the EyeSys 2000 lacks agreement with the other devices under examination across any of the parameters investigated, while both the Oculus Easygraph and ARK510A display complete interchangeability. However, both examined topographers showed perfect agreement in eccentricity measurements for everyday practice.


Acknowledgments

Funding: None.


Footnote

Data Sharing Statement: Available at https://amj.amegroups.com/article/view/10.21037/amj-23-257/dss

Peer Review File: Available at https://amj.amegroups.com/article/view/10.21037/amj-23-257/prf

Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://amj.amegroups.com/article/view/10.21037/amj-23-257/coif). The authors have no conflicts of interest to declare.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013). The study was approved by the Ethics Committee of Universidade de Santiago de Compostela (USC-40/2020) and informed consent was taken from all individual participants.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Fu D, Shang J, Zhang X, et al. Scheimpflug analysis of corneal power changes after hyperopic small incision lenticule extraction. BMC Ophthalmol 2021;21:282. [Crossref] [PubMed]
  2. Berjandy F, Nabovati P, Hashemi H, et al. Predicting initial base curve of the rigid contact lenses according to Javal keratometry findings in patients with keratoconus. Cont Lens Anterior Eye 2021;44:101340. [Crossref] [PubMed]
  3. Wu J, Fang W, Xu H, et al. The Biomechanical Response of the Cornea in Orthokeratology. Front Bioeng Biotechnol 2021;9:743745. [Crossref] [PubMed]
  4. Mehravaran S, Asgari S, Bigdeli S, et al. Keratometry with five different techniques: a study of device repeatability and inter-device agreement. Int Ophthalmol 2014;34:869-75. [Crossref] [PubMed]
  5. Hidalgo IR, Rozema JJ, Dhubhghaill SN, et al. Repeatability and inter-device agreement for three different methods of keratometry: Placido, Scheimpflug, and color LED corneal topography. J Refract Surg 2015;31:176-81. [Crossref] [PubMed]
  6. Giráldez MJ, Yebra-Pimentel E, Parafita MA, et al. Comparison of keratometric values of healthy eyes measured by javal keratometer, nidek autokeratometer, and corneal analysis system (EyeSys). International Contact Lens Clinic 2000;27:33-40. [Crossref]
  7. Amorim-de-Sousa A, Vieira AC, González-Méijome JM, et al. Age-Related Variations in Corneal Asphericity and Long-Term Changes. Eye Contact Lens 2019;45:99-104. [Crossref] [PubMed]
  8. Beiko GH, Haigis W, Steinmueller A. Distribution of corneal spherical aberration in a comprehensive ophthalmology practice and whether keratometry can predict aberration values. J Cataract Refract Surg 2007;33:848-58. [Crossref] [PubMed]
  9. Zhang YY, Jiang WJ, Teng ZE, et al. Corneal curvature radius and associated factors in Chinese children: the Shandong Children Eye Study. PLoS One 2015;10:e0117481. [Crossref] [PubMed]
  10. Liu T, Thibos LN. Compensation of corneal oblique astigmatism by internal optics: a theoretical analysis. Ophthalmic Physiol Opt 2017;37:305-16. [Crossref] [PubMed]
  11. Tzamalis A, Tsiampali C, Prousali E, et al. Reliability of Refraction, Keratometry, and Intraocular Pressure Measurements with an Automated All-in-one Device. Optom Vis Sci 2021;98:1169-76. [Crossref] [PubMed]
  12. Kumar RS, Moe CA, Kumar D, et al. Accuracy of autorefraction in an adult Indian population. PLoS One 2021;16:e0251583. [Crossref] [PubMed]
  13. Noya-Padin V, Garcia-Queiruga J, Iacubitchii M, et al. Lenstar LS900 vs EchoScan US-800: comparison between optical and ultrasound biometry with and without contact lenses and its relationship with other biometric parameters. Expert Rev Med Devices 2023;20:681-90. [Crossref] [PubMed]
  14. Armstrong RA. Statistical guidelines for the analysis of data obtained from one or both eyes. Ophthalmic Physiol Opt 2013;33:7-14. [Crossref] [PubMed]
  15. Garcia-Queiruga J, Pena-Verdeal H, Sabucedo-Villamarin B, et al. A cross-sectional study of non-modifiable and modifiable risk factors of dry eye disease states. Cont Lens Anterior Eye 2023;46:101800. [Crossref] [PubMed]
  16. Pena-Verdeal H, Noya-Padin V, Losada-Oubiña M, et al. Changes of symptomatology, tear film and ocular surface integrity one week during Somofilcon-A and Omafilcon-A lens wear. Eur J Ophthalmol 2022; Epub ahead of print. [Crossref] [PubMed]
  17. Douthwaite W, Pardhan S. Comparison of a videokeratoscope and an autokeratometer as predictors of the optimum back surface curves of rigid corneal contact lenses. Ophthalmic Physiol Opt 1997;17:409-13. [Crossref] [PubMed]
  18. Pena-Verdeal H, Garcia-Resua C, Garcia-Queiruga J, et al. Diurnal variations of tear film osmolarity on the ocular surface. Clin Exp Optom 2023;106:351-61. [Crossref] [PubMed]
  19. Armstrong RA, Davies LN, Dunne MC, et al. Statistical guidelines for clinical studies of human vision. Ophthalmic Physiol Opt 2011;31:123-36. [Crossref] [PubMed]
  20. Armstrong RA. Recommendations for analysis of repeated-measures designs: testing and correcting for sphericity and use of manova and mixed model analysis. Ophthalmic Physiol Opt 2017;37:585-93. [Crossref] [PubMed]
  21. Armstrong RA. When to use the Bonferroni correction. Ophthalmic Physiol Opt 2014;34:502-8. [Crossref] [PubMed]
  22. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-10. [Crossref] [PubMed]
  23. Carkeet A. Exact parametric confidence intervals for Bland-Altman limits of agreement. Optom Vis Sci 2015;92:e71-80. [Crossref] [PubMed]
  24. Baradaran-Rafii A, Fekri S, Rezaie M, et al. Accuracy of Different Topographic Instruments in Calculating Corneal Power after Myopic Photorefractive Keratectomy. J Ophthalmic Vis Res 2017;12:254-9. [Crossref] [PubMed]
  25. Douthwaite WA, Mallen EA. Cornea measurement comparison with Orbscan II and EyeSys videokeratoscope. Optom Vis Sci 2007;84:598-604. [Crossref] [PubMed]
  26. Dave T, Ruston D, Fowler C. Evaluation of the EyeSys model II computerized videokeratoscope. Part II: The repeatability and accuracy in measuring convex aspheric surfaces. Optom Vis Sci 1998;75:656-62. [Crossref] [PubMed]
  27. Pardhan S, Douthwaite WA. Comparison of videokeratoscope and autokeratometer measurements on ellipsoid surfaces and human corneas. J Refract Surg 1998;14:414-9. [Crossref] [PubMed]
  28. Zhao F, Wang J, Wang L, et al. An approach for simulating the fitting of rigid gas-permeable contact lenses using 3D printing technology. Cont Lens Anterior Eye 2019;42:165-9. [Crossref] [PubMed]
  29. Cho P, Lam AK, Mountford J, et al. The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting. Optom Vis Sci 2002;79:175-83. [Crossref] [PubMed]
  30. Tajbakhsh Z, Salouti R, Nowroozzadeh MH, et al. Comparison of keratometry measurements using the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer. Ophthalmic Physiol Opt 2012;32:539-46. [Crossref] [PubMed]
doi: 10.21037/amj-23-257
Cite this article as: Noya-Padin V, Pena-Verdeal H, García-Queiruga J, Nores-Palmas N, Castro-Giraldez A, Yebra-Pimentel E. Comparative analysis of anterior corneal surface characteristics measured by two topographers and two autokeratometers. AME Med J 2025;10:12.

Download Citation